A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Practical Nuclear Medicine and Utility of Phantoms for Internal Dosimetry: XCAT Compared with Zubal
Maryam Fallahpoor1, Mehrshad Abbasi1, Faraz Kalantari2
1Department of Nuclear Medicine, Vali-Asr Hospital, School of Medicine, Tehran University of Medical Sciences, Tehran 1419731351, Iran.
Purpose:
The absorbed doses for two radioisotopes, 99mTc and 131I, between previously validated Zubal phantom and the recently developed XCAT phantom were compared.
Materials And Methods:
GATE Monte Carlo code was used to simulate the statistical process of radiation. A XCAT phantom with voxel and matrix sizes similar to a standard Zubal phantom was generated. According to Medical International Radiation Dose formalism, specific absorbed fraction (SAF) values for photons and S-factors for beta particles were tabulated. The amounts of absorbed doses were calculated and compared in different organs.
Results:
The differences of gamma radiation doses, SAFs, between Zubal and XCAT are >50% in adrenal from adrenal, pancreas from pancreas and thyroid from thyroid, in lung from kidney, kidneys from lungs and in kidneys from thyroid and thyroid from kidneys. The beta radiation doses differences between Zubal and XCAT are >50% in thyroid from thyroid, bladder from bladder, kidney from kidney, liver from bladder, thyroid from bladder and kidney from thyroid. The size and distances of the organs were different between XCAT and Zubal phantoms. Denoted differences of SAF and S-factors correspond to the different organ geometries in phantoms.
Conclusion:
The results of absorbed doses in Zubal and XCAT phantoms are different. The variations prohibit easy comparison or interchangeability of dosimetry between these phantoms.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...