A Comparative Study of 3 Different Methods of Temperature Measurement in Children
Background:
The accuracy of body temperature measurement is very important in children. The accuracy and reliability of the forehead skin thermometer (FST) and infrared tympanic thermometer (TT) are inconclusive.
Objectives:
To determine: 1) the mean difference, agreement, and accuracy of body temperature detected in children by FST and ITT, as compared with the gold standard rectal mercury-in-glass thermometer (RMT); and, 2) the cut-off level for FST and ITT in the detection of fever
Material And Method:
Children ≤ 2 years old with the chief complaint of "fever" were recruited for this study. Body temperature (BT) was measured by 3 different methods, including FST 5 times; ITT 3 times; and, RMT, 1 time, for each patient. Each measurement method was performed by 3 well-trained staff members, all of whom were blinded to the recorded data of the other methods. The mean difference in body temperature, agreement by Bland-Altman method, and the verified performance of FST and ITT by diagnostic test were assessed.
Results:
A total 312 children were recruited. Body temperatures measured by FST and ITT were lower than those measured by RMT, with a mean difference of 1.04°C (p < 0.001) and 1.03°C (p < 0.001), respectively. In subgroup analysis by levels of fever the mean differences between rectal temperature and both forehead and ear temperature were statistically significantly different (p < 0.001) for all levels of fever. Regarding the diagnostic test to verify performance, the data revealed the most practical cut-off point to be 37.0°C for both FST and ITT as compared to the gold standard RMT.
Conclusion:
Both FST and ITT were found to be accurate temperature screening methods for daily clinical use. However, the cut-off points to detect fever should be lowered to 37.0°C to be consistent with gold standard measurement.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...