Correcting for H2O interference using a RAD7 electrostatic collection-based silicon detector
G De Simone1, C Lucchetti1, G Galli2
1Università "Roma Tre", Dipartimento di Scienze, Largo San Leonardo Murialdo 1, 00146 Roma, Italy.
Journal of Environmental Radioactivity
|June 3, 2016
Summary
Water molecules negatively impact electrostatic collection of polonium-218 ions on silicon detectors, reducing their efficiency. Temperature has no effect, but corrections can adjust radon readings influenced by humidity.
Area of Science:
- Environmental Science
- Nuclear Physics
- Analytical Chemistry
Background:
- Radon monitoring is crucial for assessing environmental and health risks.
- Silicon detectors are used for electrostatic collection of polonium ions, but their efficiency can be affected by environmental factors.
- Understanding the influence of water molecules on ion collection is essential for accurate radon measurements.
Purpose of the Study:
- To investigate the effect of water molecules on the electrostatic collection efficiency of polonium-218 ions onto silicon detectors.
- To compare the performance of silicon detectors with a humidity-insensitive scintillation cell under varying humidity conditions.
- To develop correction factors for soil radon readings affected by humidity and desiccant use.
Main Methods:
- Utilized a radon monitor (RAD7) connected to a stainless steel chamber with both a silicon detector and a ZnS scintillation cell.
- Manipulated water molecule concentration in the chamber by using desiccants or bubbling water.
- Measured the relative efficiency of the silicon detector against the scintillation cell across a range of water content (0.00075 to 0.014 g).
Main Results:
- Silicon detector efficiency decreased with increasing water molecules, indicating a negative impact of humidity.
- The humidity effect was more pronounced during humidification than drying, attributed to cluster formation of polonium ions.
- Temperature (18.5-35.6 °C) did not significantly affect detector efficiency.
- A humidity-dependent error range of 1.8% to 7.5% was observed.
Conclusions:
- Water molecules significantly impair the electrostatic collection efficiency of silicon detectors for polonium-218 ions.
- Developed correction factors are necessary for accurate soil radon measurements, especially after desiccant removal following potential Non-Aqueous Phase Liquids (NAPLs) leakage.
- The findings are critical for improving the reliability of radon detection systems in diverse environmental conditions.
Related Concept Videos
Atomic Absorption Spectroscopy: Interference
2.3K
Interference leads to systematic error in atomic absorption (AA) measurements by enhancing or diminishing the analytical signal or the background. These interferences can be grouped into three main categories: spectral interference, chemical interference, and physical interference.
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
2.3K
High-Performance Liquid Chromatography: Types of Detectors
2.1K
The role of the detectors in High-Performance Liquid Chromatography (HPLC) is to analyze the solutes as they exit from the chromatographic column. The detector recognizes the solute's property and generates corresponding electrical signals, which are converted into a readable graph of the detector's response versus elution time called a chromatogram at the computer. There are several types of HPLC detectors, each with its own advantages and limitations, depending on the analyte...
2.1K