Related Experiment Video
Updated: Mar 20, 2026

Inverse Probability of Treatment Weighting Propensity Score using the Military Health System Data Repository and National Death Index
Published on: January 8, 2020
Impact of the introduction of a standardised ICD programming protocol: real-world data from a single centre
Nicholas Sunderland1, Amit Kaura1, Anthony Li1
1King's College Hospital, Denmark Hill, London, UK.
Purpose:
Randomised trials have shown that empiric ICD programming, using long detection times and high detection zones, reduces device therapy in ICD recipients. However, there is less data on its effectiveness in a "real-world" setting, especially secondary prevention patients. Our aim was to evaluate the introduction of a standardised programming protocol in a real-world setting of unselected ICD recipients.
Methods:
We analysed 270 consecutive ICD recipients implanted in a single centre-135 implanted prior to protocol implementation (physician-led group) and 135 after (standardised group). The protocol included long arrhythmia detection times (30/40 or equivalent) and high rate detection zones (primary prevention lower treatment zone 200 bpm). Programming in the physician-led group was at the discretion of the implanter. The primary endpoint was time-to-any therapy (ATP or shocks). Secondary endpoints were time-to-inappropriate therapy and time-to-appropriate therapy. The safety endpoints were syncopal episodes, hospital admissions and death.
Results:
At 12 months follow-up, 47 patients had received any ICD therapy (physician-led group, n = 31 vs. standardised group, n = 16). There was a 47 % risk reduction in any device therapy (p = 0.04) and an 86 % risk reduction in inappropriate therapy (p = 0.009) in the standardised compared to the physician-led group. There was a non-significant 30 % risk reduction in appropriate therapy (p = 0.32). Results were consistent across primary and secondary prevention patients. There were no significant differences in the rates of syncope, hospitalisation, and death.
Conclusions:
In unselected patients in a real-world setting, introduction of a standardised programming protocol, using long detection times and high detection zones, significantly reduces the burden of ICD therapy without an increase in adverse outcomes.
More Related Videos
07:31Implementation of a Real-Time Psychosis Risk Detection and Alerting System Based on Electronic Health Records using CogStack
Published on: May 15, 2020
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
Related Concept Videos
Integrated Healthcare System
Standards of Care II
Standards of Care I
Health Information Technology and Healthcare Information System
Health Information Technology, commonly called HIT, integrates advanced information systems and technology in healthcare settings. Its primary functions include:
Methods of Documentation VI: Case Management Model
For example, a patient with a chronic...
Nursing Clinical Information System
A Nursing Clinical Information System (NCIS) is a specialized type of healthcare information system tailored to meet the unique needs of nursing practice. It incorporates the principles of nursing informatics to streamline information management and improve the quality of care delivery.
Critical attributes of NCIS include: