Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Common Leveling Mistakes and Errors01:17

Common Leveling Mistakes and Errors

575
A survey team is tasked with determining the elevation difference between points Point A and Point B, separated by uneven terrain. They use a leveling instrument and a leveling rod.Common MistakesMisreading the Rod: During a backsight reading at Point A, the instrumentman observes the rod partially obscured by tall grass. Instead of reading 1.135 m, they mistakenly record 1.735 m due to the misalignment of the crosshair with the wrong graduation. This error adds 0.600 m to all subsequent...
575
Accuracy and Precision01:52

Accuracy and Precision

3.1K
3.1K
Accuracy and Precision01:52

Accuracy and Precision

16.9K
Scientists typically make repeated measurements of a quantity to ensure the quality of their findings and to evaluate both the precision and the accuracy of their results. Measurements are said to be precise if they yield very similar results when repeated in the same manner. A measurement is considered accurate if it yields a result that is very close to the true or the accepted value. Precise values agree with each other; accurate values agree with a true value.  Highly accurate...
16.9K
Accuracy, limits, and approximation01:28

Accuracy, limits, and approximation

1.3K
Accuracy, limits, and approximations are common in many fields, especially in engineering calculations. These concepts are imperative for ensuring that a given value is as close as possible to its true value.
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...
1.3K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

SPINK2 in hematopoiesis and cancer: Biology and clinical implications.

Cell death discovery·2026
Same author

Evidence for the Collective Nature of Radial Flow in Pb+Pb Collisions with the ATLAS Detector.

Physical review letters·2026
Same author

Evidence for the Dimuon Decay of the Higgs Boson in pp Collisions with the ATLAS Detector.

Physical review letters·2025
Same author

Evidence for Longitudinally Polarized W Bosons in the Electroweak Production of Same-Sign W Boson Pairs in Association with Two Jets in pp Collisions at sqrt[s]=13  TeV with the ATLAS Detector.

Physical review letters·2025
Same author

Exploring plan quality: using plan complexity to quantitatively analyse the tradeoff between clinical suitability and dosimetric accuracy.

Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)·2025
Same author

Observation of tt[over ¯] Production in Pb+Pb Collisions at sqrt[s_{NN}]=5.02  TeV with the ATLAS Detector.

Physical review letters·2025

Related Experiment Video

Updated: Mar 20, 2026

Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
07:57

Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision

Published on: April 29, 2014

14.1K

Delivering RapidArc®: A comprehensive study on accuracy and long term stability.

A Scaggion1, A Negri1, M A Rossato1

  • 1Medical Physics Department, Veneto Institute of Oncology IOV-IRCCS, Padua, Italy.

Physica Medica : PM : an International Journal Devoted to the Applications of Physics to Medicine and Biology : Official Journal of the Italian Association of Biomedical Physics (AIFB)
|June 6, 2016
PubMed
Summary

This study confirms the UNIQUE Linac reliably and accurately delivers RapidArc treatments. Long-term stability and error analysis demonstrate consistent performance for Volumetric Modulated Arc Therapy (VMAT).

Keywords:
Delivery accuracyDelivery errorDynamic log filesLong-term stabilityVMAT

More Related Videos

Measurement of Dynamic Scapular Kinematics Using an Acromion Marker Cluster to Minimize Skin Movement Artifact
10:07

Measurement of Dynamic Scapular Kinematics Using an Acromion Marker Cluster to Minimize Skin Movement Artifact

Published on: February 10, 2015

20.1K
Magnetic Resonance Derived Myocardial Strain Assessment Using Feature Tracking
07:21

Magnetic Resonance Derived Myocardial Strain Assessment Using Feature Tracking

Published on: February 12, 2011

14.9K

Related Experiment Videos

Last Updated: Mar 20, 2026

Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
07:57

Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision

Published on: April 29, 2014

14.1K
Measurement of Dynamic Scapular Kinematics Using an Acromion Marker Cluster to Minimize Skin Movement Artifact
10:07

Measurement of Dynamic Scapular Kinematics Using an Acromion Marker Cluster to Minimize Skin Movement Artifact

Published on: February 10, 2015

20.1K
Magnetic Resonance Derived Myocardial Strain Assessment Using Feature Tracking
07:21

Magnetic Resonance Derived Myocardial Strain Assessment Using Feature Tracking

Published on: February 12, 2011

14.9K

Area of Science:

  • Medical Physics
  • Radiation Oncology
  • Radiotherapy Technology

Background:

  • Ensuring accuracy and reliability in radiotherapy delivery systems is crucial for effective cancer treatment.
  • Volumetric Modulated Arc Therapy (VMAT) offers precise dose conformity but requires stringent quality assurance.

Purpose of the Study:

  • To evaluate the reliability and accuracy of the UNIQUE Linac for RapidArc treatments.
  • To assess the long-term stability of the UNIQUE Linac's performance in VMAT delivery.

Main Methods:

  • Analysis of 2280 Dynalog files from 179 clinical RapidArc treatments over two years.
  • Evaluation of Multi-Leaf Collimator (MLC) motion error, gantry rotation error, and Monitor Unit (MU) delivery error.
  • Investigation into the stochastic and systematic nature of errors and their temporal variations.

Main Results:

  • Delivery errors were consistently small, suggesting stricter tolerances than TG142 recommendations.
  • MLC positional errors showed a linear relationship with leaf speed, while other VMAT parameters exhibited random behavior.
  • Gantry rotation and MU delivery accuracy were comparable to leading Linacs, with no significant variations over time.

Conclusions:

  • The UNIQUE Linac demonstrates established accuracy and reliability for RapidArc treatments.
  • The stochastic nature of delivery errors was confirmed, highlighting the system's robustness.
  • Long-term stability was confirmed, validating the UNIQUE Linac as a dependable VMAT delivery system.