Iterative reconstruction can permit the use of lower X-ray tube current in CT coronary artery calcium scoring
Related Experiment Video
Updated: Mar 19, 2026

Identifying Coronary Artery Calcification on Non-gated Computed Tomography Scans
Published on: August 28, 2018
Mark A Rodrigues1, Michelle C Williams2, Thomas Fitzgerald1
11 Department of Radiology, Royal Infirmary of Edinburgh, Edinburgh, UK.
Objective::
CT coronary artery calcium scoring (CACS) is additive to traditional risk factors for predicting future cardiac events but is associated with relatively high radiation doses. We assessed the feasibility of CACS radiation dose reduction using a lower tube current and iterative reconstruction (IR).
Methods::
Artificial noise was added to the raw data from 27 CACS studies from patients who were symptomatic to simulate lower tube current scanning (75, 50 and 25% original current). All studies were performed on the same CT scanner at 120 kVp. Data were reconstructed using filtered back projection [Quantum Denoising Software (QDS+)] and IR [adaptive iterative dose reduction three dimensional mild, standard and strong]. Agatston scores were independently measured by two readers. CACS percentile risk scores were calculated.
Results::
At 75, 50 and 25% tube currents, all adaptive iterative dose reduction (AIDR) reconstructions decreased image noise relative to QDS+ (p < 0.05). All AIDR reconstructions resulted in small reductions in Agatston score relative to QDS+ at the standard tube current (p < 0.05). Agatston scores increased with QDS+ at 75, 50 and 25% tube current (p < 0.05), whereas no significant change was observed with AIDR mild at any tested tube current. No difference in the percentile risk score with AIDR mild at any tube current occurred compared with QDS+ at standard tube current (p > 0.05). Interobserver agreement for AIDR mild remained excellent even at 25% tube current (intraclass correlation coefficient 0.997).
Conclusion::
Up to 75% reduction in CACS tube current is feasible using AIDR mild.
Advances In Knowledge::
AIDR mild IR permits low tube current CACS whilst maintaining excellent intraobserver and interobserver variability and without altering risk classification.
More Related Videos
06:57Author Spotlight: Advancing Cardiovascular Imaging - Introducing the Spatially Weighted Calcium Score for Early Disease Detection
Published on: September 22, 2023
08:02Author Spotlight: Enhanced Quantification of Cardiovascular Calcification Progression for Longitudinal Micro PET/CT Studies in Small Research Animals
Published on: November 15, 2024
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Acute Coronary Syndrome III: Diagnostic Studies
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System IV: CMRI