Related Experiment Video
Updated: Mar 19, 2026

Ultrasensitive Detection of Biomarkers by Using a Molecular Imprinting Based Capacitive Biosensor
Published on: February 16, 2018
Molecularly imprinted polymer-based bulk optode for the determination of itopride hydrochloride in physiological
F M Abdel-Haleem1, Adel Madbouly1, R M El Nashar1
1Chemistry Department, Faculty of Science, Cairo University, Gamaa Street, 12613 Giza, Egypt.
Abstract:
We report here for the first time on the use of Molecularly Imprinted Polymers as modifiers in bulk optodes, Miptode, for the determination of a pharmaceutical compound, itopride hydrochloride as an example in a concentration range of 1×10(-1)-1×10(-4)molL(-1). In comparison to the optode containing the ion exchanger only (Miptode 3), the optode containing the ion exchanger and the MIP particles (Miptode 2) showed improved selectivity over the most lipophilic species, Na(+) and K(+), by more than two orders of magnitude. For instance, the optical selectivity coefficients using Miptode 2, [Formula: see text] , were as follow: NH4(+)˂-6; Na(+)=-4.0, which were greatly enhanced in comparison with that obtained by Miptode 3. This work opens a new avenue for using miptodes for the determination of all the pharmaceutical preparations without the need for the development of new ionophores.