Related Experiment Video
Updated: Mar 19, 2026

Mouse Body Temperature Measurement Using Infrared Thermometer During Passive Systemic Anaphylaxis and Food Allergy Evaluation
Published on: September 14, 2018
Nasopharynx is well-suited for core temperature measurement during hypothermia therapy
Kotaro Hine1,2, Shigeharu Hosono1, Ken Kawabata2
1Department of Pediatrics and Child Health, Nihon University School of Medicine, Itabashi, Tokyo, Japan.
Background:
Rectal temperature is commonly used as the core temperature during therapeutic hypothermia therapy in neonates with hypoxic-ischemic encephalopathy (HIE). The purpose of this study was to examine whether nasopharyngeal temperature could serve as a substitute for rectal temperature.
Methods:
We prospectively investigated 40 neonates with HIE who underwent therapeutic hypothermia by selective head cooling, which involved cooling the body to 34°C for 72 h. During this period, nasopharyngeal temperature was measured and compared with rectal temperature every hour.
Results:
For 40 neonates included in this study, the mean rectal and nasopharyngeal temperatures were 34.3 ± 0.4°C (n = 2920) and 34.3 ± 0.4°C (n = 2920), respectively. Nasopharyngeal temperature strongly correlated with rectal temperature (R2 = 0.623, P < 0.0001) and magnitude of the mean difference between nasopharyngeal and rectal temperature varied little during the 72 h of therapeutic hypothermia.
Conclusions:
Nasopharyngeal temperature in neonates with perinatal HIE undergoing therapeutic hypothermia may be a suitable substitute for rectal temperature.
More Related Videos
06:43Esophageal Heat Transfer for Patient Temperature Control and Targeted Temperature Management
Published on: November 21, 2017
04:54Using a Combination of Indirect Calorimetry, Infrared Thermography, and Blood Glucose Levels to Measure Brown Adipose Tissue Thermogenesis in Humans
Published on: June 2, 2023
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...