Measurement of neonatal heart rate using handheld Doppler ultrasound
Amanda Dyson1, Michele Jeffrey1, Martin Kluckow1,2
1Royal North Shore Hospital, Neonatal Intensive Care Unit, Sydney, Australia.
Objective:
This pilot study aimed to determine whether handheld Doppler ultrasound is feasible and reliable for measuring neonatal heart rate (HR) when compared with ECG.
Setting:
Stable newborns were recruited from the neonatal intensive care unit and postnatal ward between July 2014 and January 2015 at Royal North Shore Hospital, Sydney, Australia.
Interventions:
Each newborn had their HR recorded every 15 s over 145 s using four different modalities: ECG, counted audible Doppler (AD) over 10 s, pulse oximetry (PO) and the Doppler display (DD).
Outcome Measures:
The correlation and variation between each modality and ECG.
Results:
51 newborns with a median gestational age of 38 weeks (27-41) and a mean weight of 2.78 kg (0.82 to 4.76) with a median postnatal age of 3 days (0-87) were studied. There was a mean difference of 0.69 bpm (95% CI -2.9 to +1.5) between AD-HR and ECG-HR with good correlation between modalities (r=0.94, p<0.01). The median time to achieve AD-HR was 3 s (1-45). The mean difference between DD-HR and ECG-HR was 5.37 bpm (95% CI -12.8 to +2.1) with moderate correlation (r=0.37, p=0.04). The mean difference between PO-HR and ECG-HR was 0.49 bpm (95% CI -1.5 to +0.51) with good correlation (r=0.99, p<0.01). The variability between AD-HR and ECG-HR decreased with decreasing weight.
Conclusions:
AD-HR correlates well with ECG-HR. Further research in the delivery room is recommended before using AD-HR in this area.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring pulse
Fetal Circulation
Two umbilical arteries transport blood from the fetus to the placenta. At the placenta, the blood absorbs oxygen and nutrients while simultaneously eliminating waste products. This oxygen-enriched and nutrient-rich blood then returns to the fetus through one...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...