[Determination of volatile organic compounds in workplace by portable gas-chromatography]
Objective:
To establish a portable gas chromatography (GC) method for the on-site rapid determination of epoxyethane, furan, dichloromethane, benzene, toluene in workplace.
Methods:
By using dynamic dilutor, the volatile organic compounds (VOCs) standard gas was prepared and drawn to the capillary column of GC (PID) for separation and analysis. Based on the retention time and peak area, the VOCs was identified and quantified.
Results:
At the conditions of 60 degrees C column temperature and 9psi pre-pressure, The five VOCs were well separated. Good linear ranges were obtained within the ranges of 0.45 - 90.72 mg/m3 for epoxyethane, 0.22 - 44.50 mg/m3 for furan, 2.19 - 437.4 mg/m3 for dichloromethane, 0.32 - 13.29 mg/m3 for benzene, 0.38 - 15.68 mg/m3 for toluene, respectively. The correlation coefficient was not less than 0.999. The detection limits were 0.03, 0.04, 0.2, 0.008 and 0.03 mg/m3, respectively. The RSD were 0.79% - 2.4%, 1.5% - 1.9%, 0.97% - 1.8%, 2.3% - 3.2% and 3.4% - 4.6%, respectively. The average recovery were 101.4% - 108.4% 85.49% - 94.98%, 97.78% - 106.2%, 99.29% - 103.5% and 95.54% - 101.7% respectively. By using the 7890A GC method, the relative tolerance for the same gas samples were 0.091% - 8.8% for epoxyethane, 3.8% - 6.7% for furan, 0.77% - 9.4% for dichloromethane, 0.24% - 1.6% for benzene, 0.31% - 8.0% for toluene.
Conclusion:
The method is portable, accurate, sensitive, rapid, and exhibits good separation and anti-interference ability. This method is suitable for the rapid detection of VOCs in workplace and also provides reference VOC detection method for the occupational health standard.
More Related Videos
05:29Profiling Volatile Compounds in Blackcurrant Fruit using Headspace Solid-Phase Microextraction Coupled to Gas Chromatography-Mass Spectrometry
Published on: June 9, 2021
05:31Gas Chromatography-Mass Spectrometry Paired with Total Vaporization Solid-Phase Microextraction as a Forensic Tool
Published on: May 25, 2021
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Volatilization