Relating pressure tuned coupled column ensembles with the solvation parameter model for tunable selectivity in gas
Khan M Sharif1, Chadin Kulsing1, Sung-Tong Chin2
1Australian Centre for Research on Separation Science, School of Chemistry, Monash University, Wellington Rd., VIC 3168, Australia.
Abstract:
The differential pressure drop of carrier gas by tuning the junction point pressure of a coupled column gas chromatographic system leads to a unique selectivity of the overall separation, which can be tested using a mixture of compounds with a wide range of polarity. This study demonstrates a pressure tuning (PT) GC system employing a microfluidic Deans switch located at the mid-point of the two capillary columns. This PT system allowed variations of inlet-outlet pressure differences of the two columns in a range of 52-17psi for the upstream column and 31-11psi for the downstream column. Peak shifting (differential migration) of compounds due to PT difference are related to a first order regression equation in a Plackett-Burman factorial study. Increased first (upstream) column pressure drop makes the second column characteristics more significant in the coupled column retention behavior, and conversely increased second (downstream) column pressure drop makes the first column characteristics more apparent; such variation can result in component swapping between polar and non-polar compounds. The coupled column system selectivity was evaluated in terms of linear solvation energy relationship (LSER) parameters, and their relation with different pressure drop effects has been constructed by applying multivariate principle component analysis (PCA). It has been found that the coupled column PT system descriptors provide a result that shows a clear clustering of different pressure settings, somewhat intermediate between those of the two commercial columns. This is equivalent to that obtained from a conventional single-column GC analysis where the interaction energy contributed from the stationary phases can be significantly adjusted by choice of midpoint PT. This result provides a foundation for pressure differentiation for selectivity enhancement.
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Types Of Column Chromatography
Gel Filtration Chromatography
When the...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...