Related Experiment Video
Updated: Mar 19, 2026

Human Fetal Blood Flow Quantification with Magnetic Resonance Imaging and Motion Compensation
Published on: January 7, 2021
Using a modified 3D-printer for mapping the magnetic field of RF coils designed for fetal and neonatal imaging
Alexander Vavoulas1, Nicholas Vaiopoulos1, Erik Hedström2
1Department of Computer Science and Biomedical Informatics, University of Thessaly, Papassiopoulou 2-4, GR 35131 Lamia, Greece.
Abstract:
An experimental setup for characterizing the magnetic field of MRI RF coils was proposed and tested. The setup consisted of a specially configured 3D-printer, a network analyzer and a mid-performance desktop PC. The setup was tested on a single loop RF coil, part of a phased array for fetal imaging. Then, the setup was used for determining the magnetic field characteristics of a high-pass birdcage coil used for neonatal MR imaging with a vertical static field. The scattering parameter S21, converted into power ratio, was used for mapping the B1 magnetic field. The experimental measurements from the loop coil were close to the theoretical results (R=0.924). A high degree of homogeneity was measured for the neonatal birdcage RF coil. The development of MR RF coils is time consuming and resource intensive. The proposed experimental setup provides an alternative method for magnetic field characterization of RF coils used in MRI.