Related Experiment Video
Updated: Mar 19, 2026

Snap Chip for Cross-reactivity-free and Spotter-free Multiplexed Sandwich Immunoassays
Published on: November 13, 2017
A reliable aptamer array prepared by repeating inkjet-spotting toward on-site measurement
Suzuyo Inoue1, Michiko Seyama1, Toru Miura1
1NTT Device Technology Laboratories, Nippon Telegraph and Telephone Corporation, 3-1 Morinosato Wakamiya, Atsugi-shi, Kanagawa 243-0198, Japan.
Abstract:
A preparation protocol is proposed for a reliable aptamer array utilizing an ink-jet spotter. We accumulated streptavidin and biotinylated-aptamer in this order on a biotinylated-polyethylene glycol-coated gold substrate to prepare an aptamer array. The aptamer array was prepared with an alternate spotting structure where each aptamer spot was placed between reference spots formed with blocking solution thus suppressing contamination from neighboring spots during the blocking and washing processes. Four aptamer spots were prepared in a small area of 1×4.8mm(2) with five reference spots made of blocking solution. We evaluated the thrombin binding ability of the spotted aptamer array using a multi-analysis surface plasmon resonance sensor. We prepared a disposable capillary-driven flow chip designed for on-site measurement (Miura et al., 2010) with our aptamer array and detected thrombin from phosphate-buffered saline at concentrations of 50ngmL(-1) and 1μgmL(-1) (equivalent to 1.35 and 27nM, respectively). A correlation was observed between the refractive index shift and thrombin concentration. This implies that our array preparation protocol meets the requirement for the preparation of a one-time-use chip for on-site measurement.

