Improve the image quality of orbital 3 T diffusion-weighted magnetic resonance imaging with readout-segmented
Xiao-Quan Xu1, Jun Liu1, Hao Hu1
1Department of Radiology, The First Affiliated Hospital of Nanjing, Medical University, Nanjing, China.
Abstract:
The aim of our study is to compare the image quality of readout-segmented echo-planar imaging (rs-EPI) and that of standard single-shot EPI (ss-EPI) in orbital 3 T diffusion-weighted (DW) magnetic resonance (MR) imaging in healthy subjects. Forty-two volunteers underwent two sets of orbital DW imaging scan at a 3 T MR unit, and image quality was assessed qualitatively and quantitatively. As a result, we found that rs-EPI could provide better image quality than standard ss-EPI, while no significant difference was found on the apparent diffusion coefficient between the two sets of DW images.