Improved procedure for dendrimer-based mass calibration in matrix-assisted laser
1Institute of Organic Chemistry, Heidelberg University, Im Neuenheimer Feld 270, 69120, Heidelberg, Germany. juergen.gross@oci.uni-heidelberg.de.
Abstract:
A procedure is described that results in a substantial increase in signal intensity and in improved accuracy of positive-ion mass calibration when using commercially available kits of monodisperse dendrimers (SpheriCal(®)) in matrix-assisted laser desorption/ionization (MALDI) mass spectrometry (MS). The peak intensities are amplified by an admixture of 2-[(2E)-3-(4-tert-butylphenyl)-2-methylprop-2-enylidene] malononitrile (DCTB) matrix to the kits comprising of 9-nitroanthracene matrix, sodium trifluoroacetate, and four dendrimers. Boosted ion formation then permits lower laser fluence to be used and thus yields enhanced mass resolution. Further, the number of reference peaks is doubled by doping the sample preparation with cesium ions. This results in four [M+Cs](+) ion signals in addition to four [M+Na](+) ion signals provided by the standard kit. Overall, the modified procedure notably reduces the consumption of the expensive calibration standard kits, while it increases mass resolution and enables the use of an advanced calibration algorithm requiring at least six reference peaks. Graphical abstract A dendrimer-based mass calibration for MALDI-TOF-MS can be improved by adding a DCTB matrix and doping the sample preparation with Cs(+) ions. Having eight rather than just four reference peaks reduces the average mass error of the calibration curve about fivefold.
More Related Videos
07:49Matrix-assisted Laser Desorption/Ionization Time of Flight MALDI-TOF Mass Spectrometric Analysis of Intact Proteins Larger than 100 kDa
Published on: September 9, 2013
09:38Dithranol as a Matrix for Matrix Assisted Laser Desorption/Ionization Imaging on a Fourier Transform Ion Cyclotron Resonance Mass Spectrometer
Published on: November 26, 2013
Related Concept Videos
Matrix-Assisted Laser Desorption Ionization (MALDI)
MALDI-TOF Mass Spectrometry
Tandem Mass Spectrometry
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...