Measuring the Carotid to Femoral Pulse Wave Velocity Cf-PWV to Evaluate Arterial Stiffness
Published on: May 3, 2018
Comparison of Central Blood Pressure Estimated by a Cuff-Based Device With Radial Tonometry
Xiaoqing Peng1, Martin G Schultz1, Walter P Abhayaratna2
1Menzies Institute for Medical Research, University of Tasmania, Hobart, Australia;
Background:
New techniques that measure central blood pressure (BP) using an upper arm cuff-based approach require performance assessment. The aim of this study was to compare a cuff-based device (CuffCBP) to estimate central BP indices (systolic BP (SBP), diastolic BP (DBP), pulse pressure (PP), augmentation pressure (AP), augmentation index (AIx)) with noninvasive radial tonometry (TonCBP).
Methods:
Consecutive CuffCBP (SphygmoCor Xcel) and TonCBP (SphygmoCor 8.1) duplicate recordings were measured in 182 people with treated hypertension (aged 61±7 years, 48% male). Agreement between methods was assessed using standard calibration with brachial SBP and DBP (measured with the Xcel device), as well as with brachial mean arterial pressure (MAP; 40% form factor method) and DBP.
Results:
The mean difference ± SD for central SBP (cSBP), central DBP (cDBP), and central PP (cPP) between methods were -0.89±3.48mm Hg (intra-class correlation (ICC) 0.977; 95% confidence interval (CI) 0.973-0.982), -0.50±1.54mm Hg (ICC 0.992, 95% CI 0.987-0.993), and -0.42±3.57mm Hg (ICC 0.966, 95% CI 0.958-0.972), indicating good agreement. Wider limits of agreement were observed for central AP (cAP) and central AIx (cAIx) (-0.91±5.31mm Hg; ICC 0.802; 95% CI 0.756-0.839, -0.99±10.91%; ICC 0.749; 95% CI 0.691-0.796). Re-calibration with brachial MAP and DBP resulted in an overestimation of cSBP with CuffCBP compared with TonCBP (8.58±19.06mm Hg, ICC 0.164, 95% CI -0.029 to 0.321).
Conclusion:
cSBP, cDBP, and cPP derived from CuffCBP are substantially equivalent to TonCBP, although the level of agreement is dependent on calibration method. Further validity testing of CuffCBP by comparison with invasively measured central BP will be required.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: