Cardiac Output Assessed by the Fourth-Generation Arterial Waveform Analysis System Is Unreliable in Liver Transplant
B-F Shih1, P-H Huang1, H-P Yu2
1Department of Anesthesiology, Chang Gung Memorial Hospital, Guishan, Taoyuan, Taiwan.
Background:
Liver transplant recipients often have violent hemodynamic fluctuation during surgery that may be related to perioperative and postoperative morbidity. Because there are some considerations for the risk of the pulmonary arterial catheter (PAC), the conventional invasive device for cardiac output (CO) measurement, a reliable and minimally invasive alternative is required. We validated the reliability of CO measurements with the use of a minimally invasive FloTrac system with the latest fourth-generation algorithm in liver transplant recipients.
Methods:
Forty liver transplant recipients without atrial fibrillation, valvular pathology, or intracardiac shunt were recruited in this prospective, observational study. CO values measured by use of PAC with continuous thermodilution method (COTh) and FloTrac devices (COFT) were collected simultaneously throughout the operation for reliability validation.
Results:
Four hundred pairs of CO data points were collected in total. The linear regression analysis showed a high correlation coefficient (73%, P < .001). However, the percent error between COTh and COFT was 42.2%, which is worse than the established interchangeability criterion of 30%. The concordance rates were calculated at 89% and 59% by 4-quadrant plot and polar plot analysis, respectively. Neither met the preset validation criteria (>92% for the 4-quadrant plot and >90% for polar plot analyses).
Conclusions:
Our study demonstrates that the CO measurements in liver transplant recipients by the latest FloTrac system and the PAC do not meet the recognized interchangeability criterion. Although the result showed improvement in linear regression analysis, it failed to display a qualified trending ability.
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...