Missing LINQ: extrusion of a new-generation implantable loop recorder in a child
Related Experiment Video
Updated: Oct 12, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Ahmad S Chaouki1, Richard J Czosek1, David S Spar1
1Department of Pediatrics,Division of Cardiology,The Heart Institute,Cincinnati Children's Hospital Medical Center,University of Cincinnati College of Medicine,Cincinnati,Ohio,United States of America.
Abstract:
Cardiac rhythm monitoring has been facilitated by the use of implantable loop recorders. New models of these devices are 87% smaller than before allowing for easier implantation and use in the paediatric population. Recommendations are for closure with adhesive. We report a device extrusion in a 6-year-old patient. Based on this, our practice has changed to include subcutaneous sutures this complication.