Periprosthetic Artifact Reduction Using Virtual Monochromatic Imaging Derived From Gemstone Dual-Energy Computed
Exequiel Reynoso1, Carlos Capunay, Alejandro Rasumoff
1From the Department of Computed Tomography, Diagnóstico Maipú, Buenos Aires, Argentina.
Objective:
The aim of this study was to explore the usefulness of combined virtual monochromatic imaging and metal artifact reduction software (MARS) for the evaluation of musculoskeletal periprosthetic tissue.
Methods:
Measurements were performed in periprosthetic and remote regions in 80 patients using a high-definition scanner. Polychromatic images with and without MARS and virtual monochromatic images were obtained.
Results:
Periprosthetic polychromatic imaging (PI) showed significant differences compared with remote areas among the 3 tissues explored (P < 0.0001). No significant differences were observed between periprosthetic and remote tissues using monochromatic imaging with MARS (P = 0.053 bone, P = 0.32 soft tissue, and P = 0.13 fat). However, such differences were significant using PI with MARS among bone (P = 0.005) and fat (P = 0.02) tissues. All periprosthetic areas were noninterpretable using PI, compared with 11 (9%) using monochromatic imaging.
Conclusions:
The combined use of virtual monochromatic imaging and MARS reduced periprosthetic artifacts, achieving attenuation levels comparable to implant-free tissue.