Comparison between visual score and erythema index (DermaSpectrometer) in evaluation of allergic patch tests.
E Held1, H Lorentzen1, T Agner1
1Department of Dermatology, University of Copenhagen, Gentofte HospitalCopenhagen Wound Healing Center, University of Copenhagen, Bispebjerg Hospital, Copenhagen, Denmark.
Summary
The DermaSpectrometer
Area of Science:
- Dermatology
- Allergology
- Bioengineering
Background:
- Objective evaluation of patch test reactions is an ongoing area of bioengineering research.
- The erythema index, measured by the DermaSpectrometer, offers a potential objective method for assessing patch test responses.
- Allergic contact dermatitis diagnosis relies heavily on accurate patch test interpretation.
Purpose of the Study:
- To evaluate the clinical utility of the DermaSpectrometer for measuring erythema index in allergic patch tests.
- To compare DermaSpectrometer readings with traditional visual assessments of patch test reactions.
- To determine if the DermaSpectrometer provides more objective data than visual inspection in formaldehyde patch testing.
Main Methods:
- Twenty patients with known formaldehyde allergy underwent patch testing with varying formaldehyde concentrations.
- Erythema index was measured using the DermaSpectrometer 24 hours after patch test removal.
- Clinical readings of patch test sites were performed concurrently, and a control group was included.
Main Results:
- Significantly higher erythema indices were observed in visually positive patch tests compared to negative ones (P<0.05).
- The DermaSpectrometer could not reliably differentiate between various grades of visually positive reactions.
- Correlation between visual clinical readings and formaldehyde concentration (r=0.60) was stronger than DermaSpectrometer readings (r=0.35).
Conclusions:
- The DermaSpectrometer did not provide superior information compared to visual readings in formaldehyde patch testing.
- Visual assessment remains a valuable method for interpreting patch test reactions, even with the availability of objective tools.
- Further research may be needed to refine bioengineering tools for more accurate patch test evaluation.