Increased dead space in face mask continuous positive airway pressure in neonates
Kenji Hishikawa1, Hideshi Fujinaga1, Yushi Ito1
1Division of Neonatology, Center of Maternal-Fetal, Neonatal and Reproductive Medicine, National Center for Child Health and Development, Tokyo, Japan.
The Objectives:
Continuous positive airway pressure (CPAP) by face mask is commonly performed in newborn resuscitation. We evaluated the effect of face mask CPAP on system dead space.
Working Hypothesis:
Face mask CPAP increases dead space.
Study Design:
A CPAP model study.
Methodology:
We estimated the volume of the inner space of the mask. We devised a face mask CPAP model, in which the outlet of the mask was covered with plastic; and three modified face mask CPAP models, in which holes were drilled near to the cushion of the covered face mask to alter the air exit. We passed a continuous flow of 21% oxygen through each model and we controlled the inner pressure to 5 cmH2 O by adjusting the flow-relief valve. To evaluate the ventilation in the inner space of each model, we measured the oxygen concentration rise time, that is, the time needed for the oxygen concentration of each model to reach 35% after the oxygen concentration of the continuous flow was raised from 21% to 40%.
Results:
The volume of inner space of the face mask was 38.3 ml. Oxygen concentration rise time in the face mask CPAP model was significantly longer at various continuous flow rates and points of the inner space of the face mask compared with that of the modified face mask CPAP model.
Conclusions:
Our study indicates that face mask CPAP leads to an increase in dead space and a decrease in ventilation efficiency under certain circumstances. Pediatr Pulmonol. 2017;52:107-111. © 2016 Wiley Periodicals, Inc.
More Related Videos
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
07:52Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...