Related Experiment Video
Updated: Mar 19, 2026

Use of Electromagnetic Navigational Transthoracic Needle Aspiration E-TTNA for Sampling of Lung Nodules
Published on: May 23, 2015
Flat detector cone-beam CT-guided percutaneous needle biopsy of mediastinal lesions: preliminary experience
Dechao Jiao1, Kai Huang2, Gang Wu1
1Department of Interventional Radiology, The First Affiliated Hospital of Zhengzhou University, Zhengzhou, 450052, Henan, People's Republic of China.
Purpose:
The purpose of this study was to evaluate the usefulness of flat detector cone-beam CT-guided CBCT percutaneous needle biopsy (PNB) of mediastinal lesions.
Methods:
A total of 100 patients with 100 solid mediastinal lesions were retrospectively enrolled to undergo percutaneous needle biopsy (PNB) procedures. The mean diameter of lesions was 4.4 ± 1.8 cm (range 1.8-9.0 cm). The needle path was carefully planned and calculated on the CBCT virtual navigation guidance system, which acquired 3D CT-like cross-sectional images. Diagnostic performance, procedure details, complication rate, and patient radiation exposure were investigated.
Results:
The technical success rate of PNB under CBCT virtual navigation system was 100 % (100/100). The sensitivity, specificity, and accuracy of PNB of small nodules under iGuide CBCT virtual navigation guidance were 95.1 % (79/83), 100 % (12/12), and 95.7 % (91/95), respectively. The number of biopsies and CBCT acquisitions were 2.6 ± 1.2 (range 1-6) and 3.0 ± 1.1 (range 2-8), respectively. Complications occurred in five (5.0 %) cases. The mean total procedure time was 11.70 ± 3.44 min (range 6-27 min), resulting in a mean exposure dose of 9.7 ± 4.3 mSv.
Conclusion:
Flat detector cone-beam CT-guided PNB is an accurate and safe diagnostic method for mediastinal lesions.