Effects of Cobalt and Chromium Levels Following Modular Hip Stem Total Hip Arthroplasty

Orthopedics
|June 24, 2016
PubMed

Insights

Revision total hip arthroplasty (THA) decisions for modular hip stem fretting rely solely on patient pain levels. Objective markers like serum cobalt, chromium, or MARS MRI findings did not reliably predict the need for revision THA.

Area of Science:

  • Orthopedic Surgery
  • Biomaterials Science
  • Diagnostic Imaging

Background:

  • Increasing rates of revision total hip arthroplasties (THAs) are linked to pain from modular hip stem fretting.
  • Objective markers are sought to identify patients requiring revision THA due to fretting.

Purpose of the Study:

  • To correlate serum cobalt (Co) and chromium (Cr) levels, and MARS MRI findings with the need for revision THA.
  • To evaluate the utility of objective diagnostic tools in assessing fretting-related pain in THA patients.

Main Methods:

  • Retrospective study of 66 patients undergoing THA.
  • Measured serum Co and Cr levels, and utilized MARS MRI in a subset of patients.
  • Correlated objective findings with patient-reported pain scores and need for revision surgery.

Main Results:

  • Serum Cr levels showed a trend towards higher levels in patients with positive MARS MRI findings (P=.083).
  • Serum Co levels approached significance in patients undergoing revision THA (P=.076).
  • Patient pain scores were significantly higher in the revision THA group (P<.001).

Conclusions:

  • No objective findings (serum metal ions or MARS MRI) reliably identified patients needing revision THA for fretting.
  • Patient-reported pain remains the primary factor in deciding on revision THA for modular hip stem fretting.