Total mesorectal excision using a soft and flexible robotic arm: a feasibility study in cadaver models
Alberto Arezzo1, Yoav Mintz2, Marco Ettore Allaix3
1Department of Surgical Sciences, University of Torino, Corso Dogliotti 14, 10126, Turin, Italy. alberto.arezzo@unito.it.
Surgical Endoscopy
|June 25, 2016
Summary
The STIFF-FLOP robotic optic arm, a flexible device, successfully assisted in laparoscopic total mesorectal excision (TME) in cadavers. This soft robotic technology offers a superior field of vision, potentially overcoming limitations of rigid instruments in complex surgeries.
Area of Science:
- Robotics in Surgery
- Minimally Invasive Procedures
- Surgical Technology Innovation
Background:
- The STIFF-FLOP project developed a novel stiffness-controllable flexible manipulator for surgical applications.
- This research addresses limitations of current rigid-link robotic systems in surgery.
- The study introduces a soft, flexible robotic arm for enhanced optic vision during procedures.
Purpose of the Study:
- To evaluate the feasibility and effectiveness of the STIFF-FLOP robotic optic arm in cadaveric total mesorectal excision (TME).
- To assess the visual field advantages of the flexible robotic optic compared to standard laparoscopic vision.
- To explore the potential of soft robotics in overcoming challenges in difficult laparoscopic surgeries.
Main Methods:
- The STIFF-FLOP robotic optics system was utilized in two embalmed male cadavers for TME procedures.
- The flexible optic prototype comprised two 60 mm modules with a 14.3 mm maximum outer diameter.
- Procedures involved standard laparoscopic instruments and a 6-degree-of-freedom manipulator arm controlled via joysticks.
Main Results:
- Laparoscopic TME was successfully completed in both cadavers using the STIFF-FLOP robotic optics.
- The flexible robotic arm provided superior angles of vision in the pelvis, ensuring an intact mesorectum.
- Procedure times were 165 and 145 minutes, with no intraoperative complications or technical failures recorded.
Conclusions:
- The STIFF-FLOP soft and flexible robotic optic arm is effective for assisting laparoscopic TME in a cadaver model.
- It offers a superior field of vision, particularly in the deep pelvis, compared to standard laparoscopy.
- Soft and flexible robotic devices show promise for improving outcomes in challenging laparoscopic surgeries.