Implementation of a patient dose monitoring system in conventional digital X-ray imaging: initial experiences
Christina Heilmaier1, Niklaus Zuber2, Dominik Weishaupt2
1Department of Radiology and Nuclear Medicine, Stadtspital Triemli Zurich, Birmensdorferstrasse 497, CH-8063, Zurich, Switzerland. christina.heilmaier@triemli.zuerich.ch.
Objectives:
The purpose was to report on the initial experience after implementation of a patient dose-monitoring system in conventional X-ray imaging.
Methods:
A dose-monitoring system collected dose data relating to different radiographs (one projection) and studies (two or more projections). Images were acquired on digital X-ray systems equipped with flat-panel detectors. During period 1, examinations were performed in a routine fashion in 12,614 patients. After period 1, technical modifications were performed and radiographers underwent training in radiation protection. During period 2, examinations were performed in 14,514 patients, and the radiographers were advised to read dose data after each radiograph/study. Dose data were compared by means of kerma area product (KAP, gray × centimetre squared) and entrance surface air kerma (ESAK, milligray).
Results:
During period 1, 13,955 radiographs and 8,466 studies were performed, and in period 2 16,090 radiographs and 10,389 studies. In period 2, KAP values for radiographs were an average of 25 % lower and for studies 7 % lower, and ESAK values for radiographs were 24 % lower and for studies 5 % lower. The reduction in KAP was significant in 8/13 radiographs and in 6/14 studies, and the reduction in ESAK was significant in 6/13 radiographs and 5/14 studies.
Conclusions:
Implementation of a patient dose-monitoring system in conventional X-ray imaging allows easy data collection, supports dose reduction efforts, and may increase radiographers' dose awareness.
Key Points:
• A dose-monitoring system enables easy dose data collection in X-ray imaging. • The system facilitates dose reduction efforts and may increase radiographers' radiation awareness. • Mean kerma-area-product significantly declined in period 2 in 8/13 radiographs and 6/14 studies. • In ESAK a significant decline was evident in 6/13 radiographs and 5/14 studies.
More Related Videos
Related Concept Videos
X-ray Imaging
Therapeutic Drug Monitoring: Drug Analysis Methods
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Therapeutic Drug Monitoring: Overview and Classification
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT