Related Experiment Video
Updated: Mar 18, 2026

Low-Dose Gamma Radiation Sterilization for Decellularized Tracheal Grafts
Published on: April 14, 2023
Microbiological Load of Ethylene Oxide Sterilized Medical Devices and its Elimination by Cobalt 60 Source
Rasheeda Bashir1, Bakht Afroze1, Hafiza Fizza Zulfiqar1
1Department of Biotechnology, Lahore College for Women University, Lahore.
Objective:
To determine the residing microbial flora of ethylene oxide (EtO) sterilized medical devices and optimization of safe dose of gamma radiation (Cobalt 60 source) for the complete elimination of microbial load.
Study Design:
Experimental study.
Place And Duration Of Study:
Department of Biotechnology, Lahore College for Women University, Lahore, Pakistan from September 2014 to June 2015.
Methodology:
Thirty-six samples of EtO sterilized medical devices of same batch of three different companies were collected for this study. Isolation and enumeration of microbes were done by using different selective and differential media. Gram staining and biochemically characterization by API 20 (Bio Merieux, France) kit was done for identification of the microorganisms. The medical devices having high microbial load were sent to Pakistan Radiation Services (PARAS) for gamma irradiations at 3 different selected doses (20 KGy, 25 KGy, and 30 KGy).
Results:
Different types of Gram positive bacteria (Staphylococcus epidermidis, Staphylococcus aureus andBacillus subtilis) were isolated from the EtO sterilized samples. Gram negative bacteria and fungi were not detected on these medical devices. Gamma irradiations results showed that 30 KGy was optimized dose for complete elimination of microbial flora on endotracheal, Nelaton, and tracheostomy tubes.
Conclusion:
Gamma radiations (Co 60 source) effectively decontaminate the microbial flora on the equipment previously sterilized by the ethylene oxide gas; and 30 KGy is the optimized dose for all these medical devices.
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Chemical Agents for Microbial Control
Microbial Bioremediation of Uranium