Assessment of sub-milli-sievert abdominal computed tomography with iterative reconstruction techniques of different
Atul Padole1, Nisha Sainani1, Diego Lira1
1Atul Padole, Nisha Sainani, Diego Lira, Ranish Deedar Ali Khawaja, Sarvenaz Pourjabbar, Roberto Lo Gullo, Alexi Otrakji, Mannudeep K Kalra, Department of Radiology, Massachusetts General Hospital, Boston, MA 02114, United States.
Aim:
To assess diagnostic image quality of reduced dose (RD) abdominal computed tomography (CT) with 9 iterative reconstruction techniques (IRTs) from 4 different vendors to the standard of care (SD) CT.
Methods:
In an Institutional Review Board approved study, 66 patients (mean age 60 ± 13 years, 44 men, and 22 women) undergoing routine abdomen CT on multi-detector CT (MDCT) scanners from vendors A, B, and C (≥ 64 row CT scanners) (22 patients each) gave written informed consent for acquisition of an additional RD CT series. Sinogram data of RD CT was reconstructed with two vendor-specific and a vendor-neutral IRTs (A-1, A-2, A-3; B-1, B-2, B-3; and C-1, C-2, C-3) and SD CT series with filtered back projection. Subjective image evaluation was performed by two radiologists for each SD and RD CT series blinded and independently. All RD CT series (198) were assessed first followed by SD CT series (66). Objective image noise was measured for SD and RD CT series. Data were analyzed by Wilcoxon signed rank, kappa, and analysis of variance tests.
Results:
There were 13/50, 18/57 and 9/40 missed lesions (size 2-7 mm) on RD CT for vendor A, B, and C, respectively. Missed lesions includes liver cysts, kidney cysts and stone, gall stone, fatty liver, and pancreatitis. There were also 5, 4, and 4 pseudo lesions (size 2-3 mm) on RD CT for vendor A, B, and C, respectively. Lesions conspicuity was sufficient for clinical diagnostic performance for 6/24 (RD-A-1), 10/24 (RD-A-2), and 7/24 (RD-A-3) lesions for vendor A; 5/26 (RD-B-1), 6/26 (RD-B-2), and 7/26 (RD-B-3) lesions for vendor B; and 4/20 (RD-C-1) 6/20 (RD-C-2), and 10/20 (RD-C-3) lesions for vendor C (P = 0.9). Mean objective image noise in liver was significantly lower for RD A-1 compared to both RD A-2 and RD A-3 images (P < 0.001). Similarly, mean objective image noise lower for RD B-2 (compared to RD B-1, RD B-3) and RD C-3 (compared to RD C-1 and C-2) (P = 0.016).
Conclusion:
Regardless of IRTs and MDCT vendors, abdominal CT acquired at mean CT dose index volume 1.3 mGy is not sufficient to retain clinical diagnostic performance.
More Related Videos
05:32Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...