Endotracheal Tube Cuff Pressure Monitoring in Peripheral Hospitals
R Goyal1, G Kumar2, M R Waghray3
1Graded Specialist (Anaesthesia), 164 MH C/o 99 APO., New Delhi.
Background:
An improvised monitor was designed in a peripheral hospital to measure the tracheal tube cuff pressures in patients intubated under anaesthesia. The aim of the study was to assess the efficacy of assessment of cuff pressure by the traditional palpatory method and to compare the improvised monitor with the standard monitor commercially available. The effect of nitrous oxide on the cuff-pressure was also studied.
Methods:
The tracheal tube cuff pressure of 80 patients undergoing general anaesthesia was assessed by palpation and measured with an improvised and standard monitor.
Results:
The study showed that the tracheal cuff pressure recorded were higher than normal tracheal perfusion pressure in 40% of the cases with satisfactory palpatory assessment. The pressures recorded by the improvised monitor were comparable to that of the standard monitor. The use of nitrous oxide resulted in increase in cuff pressures over a period of time.
Conclusion:
An objective measurement by any equipment is superior to assessment of cuff pressure by palpation. The improvised monitor can be used to give a fair idea of the cuff pressures, in places where a standard monitor is not available.
More Related Videos
04:20Integration of Brain Tissue Saturation Monitoring in Cardiopulmonary Exercise Testing in Patients with Heart Failure
Published on: October 1, 2019
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...