Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Photopic and Mesopic Performance of 2 Different Trifocal Diffractive Intraocular Lenses
Jose M Martinez-de-la-Casa1,2,3, Jesús Carballo-Alvarez4, Javier Garcia-Bella1,2,3
11 Department of Ophthalmology, Hospital Clínico San Carlos, Madrid - Spain.
Abstract:
Purpose To assess photopic and mesopic vision in patients implanted with 2 different trifocal intraocular lenses (IOLs). Methods Fifty patients with cataract in both eyes had surgery with bilateral implantation of a trifocal FineVision (FV, n = 25) or AT LISA tri 839MP (LisaTri, n = 25) IOL. Three months after surgery, high-contrast photopic uncorrected distance visual acuity and corrected distance visual acuity (CDVA) were determined using the Early Treatment Diabetic Retinopathy Study test. Intermediate (65 cm) and near (40 cm) visual acuity were also measured, both with best distance correction (DCIVA and DCNVA, respectively). The CSV1000 test chart was used for binocular and monocular contrast sensitivity. Defocus curves were constructed under photopic and mesopic conditions. Results Mean patient age was 68.4 ± 5.5 years (FV) and 67.3 ± 5.6 years (LisaTri) (p = 0.540). No differences were found in terms of visual acuity between the groups. The CDVA, DCIVA, and DCNVA were 0.05 ± 0.06, 0.25 ± 0.10, and 0.13 ± 0.10 logMAR for FV and -0.04 ± 0.11, 0.32 ± 0.09, and 0.12 ± 0.11 logMAR for LisaTri, respectively (p = 0.09, p = 0.08, and p = 0.12). Photopic and mesopic defocus curves were comparable between the 2 IOL groups (p>0.05). Mesopic values were significantly worse than photopic in both groups for all the measured values (p<0.001). No differences in contrast sensitivity were observed between the 2 IOL groups (p>0.05). Conclusions Both trifocal diffractive IOLs provided good visual quality with a statistically significant decrease in mesopic conditions.
More Related Videos
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
05:46Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024