Related Experiment Video
Updated: Mar 18, 2026

06:57
Leukodepletion Filters-Derived CD34+ Cells As a Cell Source to Study Megakaryocyte Differentiation and Platelet Formation
Published on: May 20, 2021
4.6K
Comparative Study of Predeposit and Bedside Leucodepletion Filters.
A Sen1, A Khetarpal Sm2, S Jetley3
1Classified Specialist (Pathology & Micro), Army Hospital (R&R), Delhi Cantt-10.
Medical Journal, Armed Forces India
|July 5, 2016
Summary
Prestorage leucodepletion using polyurethane filters is more effective than bedside methods for packed red blood cells (RBCs), ensuring better red cell recovery and higher leucocyte reduction. This method is recommended for specific patient needs.
Area of Science:
- Blood banking and transfusion medicine
- Cellular therapy and component processing
Background:
- Leukocyte removal from cellular blood components reduces transfusion-associated adverse reactions.
- Optimizing leukocyte reduction is crucial for transfusion safety.
Purpose of the Study:
- To compare the efficacy of prestorage versus bedside leucodepletion of packed red blood cells (RBCs).
- To evaluate different filter types (polyurethane vs. polyester) and temperatures for leucodepletion.
- To determine the optimal strategy for achieving log 3 leucoreduction.
Main Methods:
- 400 units of packed RBCs underwent leucodepletion using prestorage and bedside filters (Terumo Penpol Immugard III, Pall Medical BPF-4) at room temperature and 4°C.
- Pre- and post-filtration parameters, including red cell recovery and residual leukocyte count, were analyzed.
Main Results:
- All tested filters achieved over log 3 leucoreduction.
- Prestorage leucodepletion with polyurethane filters yielded higher leucocyte reduction than polyester filters.
- Bedside filters at room temperature showed significantly lower red cell recovery compared to prestorage filters.
Conclusions:
- Prestorage leucodepletion is superior to bedside leucodepletion for packed RBCs.
- Polyurethane filters are more effective than polyester filters for leucodepletion.
- Selective log 3 leucodepletion using polyurethane prestorage filters is recommended for specific patient indications.