Dose prescription in carbon ion radiotherapy: How to compare two different RBE-weighted dose calculation systems
Silvia Molinelli1, Giuseppe Magro1, Andrea Mairani2
1CNAO Foundation, Pavia, Italy.
Background And Purpose:
In carbon ion radiotherapy (CIRT), the use of different relative biological effectiveness (RBE) models in the RBE-weighted dose (DRBE) calculation can lead to deviations in the physical dose (Dphy) delivered to the patient. Our aim is to reduce target Dphy deviations by converting prescription dose values.
Material And Methods:
Planning data of patients treated at the National Institute of Radiological Sciences (NIRS) were collected, with prescribed doses per fraction ranging from 3.6Gy (RBE) to 4.6Gy (RBE), according to the Japanese semi-empirical model. The Dphy was Monte Carlo (MC) re-calculated simulating the NIRS beamline. The local effect model (LEM)_I was then applied to estimate DRBE. Target median DRBE ratios between MC+LEM_I and NIRS plans determined correction factors for the conversion of prescription doses. Plans were re-optimized in a LEM_I-based commercial system, prescribing the NIRS uncorrected and corrected DRBE.
Results:
The MC+LEM_I target median DRBE was respectively 15% and 5% higher than the NIRS reference, for the lowest and highest dose levels. Uncorrected DRBE prescription resulted in significantly lower target Dphy in re-optimized plans, with respect to NIRS plans.
Conclusions:
Prescription dose conversion factors could minimize target physical dose variations due to the use of different radiobiological models in the calculation of CIRT RBE-weighted dose.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Biological Effects of Radiation
Dose Size and Dosing Frequency: Determination Methods
Radiation: Applications
The average...
Isotopes and Radioisotopes
An isotope containing...