Failure of tension band plating: a case series
Julio J Masquijo1, Gregory B Firth, Dalia Sepúlveda
1aDepartment of Pediatric Orthopaedics, Sanatorio Allende, Córdoba, Argentina bDepartment of Orthopaedic Surgery, Chris Hani Baragwanath Academic Hospital, University of the Witwatersrand, South Africa cCOTI, Santiago, Chile.
Abstract:
Growth modulation with tension band plates (TBP) has been shown to be a very useful method for the treatment of angular deformities in growing children. Recently, we have observed cases of failure where the epiphyseal screw was drawn through the physis into the metaphysis. This study describes a series of children who developed this complication. Patients who developed TBP failure after operative treatment of lower limb angular deformities were identified from the databases at four institutions over a 5-year period. The medical records were reviewed to record demographics, primary diagnoses, details of the operative procedure, development of physeal arrest, and recurrence of the original deformity. Six patients (five girls) with nine implant failures were identified. The mean age of the children at the time of implant insertion was 7.2 years (range, 4-10 years). The primary diagnoses included hypophosphatemic rickets (n=7), congenital pseudoarthrosis of the tibia associated with neurofibromatosis 1 (n=1), and post-traumatic malunion after distal tibial fracture (n=1). Of the nine TBP that presented with the complication, four were inserted into the medial distal femur (one bilateral case), two into the medial proximal tibia (one bilateral case), two into the lateral distal tibia, and one into the medial distal tibia. None of these patients developed physeal growth arrest at the last follow-up as assessed on the latest radiographs. The use of TBP for guided growth in patients younger than 10 years old with rickets, neurofibromatosis, or other conditions that produce osteopenia leads to an increased risk for implant failure. In these cases, it is important to confirm that the epiphyseal screw has good purchase. Patients with these features should be monitored closely for early detection of this complication.
Related Concept Videos
Series R—L Circuit Transients
Using Kirchhoff's Voltage Law (KVL) to analyze this circuit helps determine the total asymmetrical fault current, which consists...
Cable: Problem Solving
RLC Series Circuit: Problem-Solving
To solve the problem, first, determine the known and unknown quantities in the problem. Recalling the reactance equation for the inductor and capacitor and substituting the values,...
Resistors In Series
In a series circuit, the...
Magnetic Force On Current-Carrying Wires: Example
Cable Subjected to Its Own Weight
A generalized loading function is employed to analyze a cable subjected to its own weight. This function considers the force acting along the cable's arc length rather than its projected length, providing a more accurate...