The Diagnostic Concordance of Whole Slide Imaging and Light Microscopy: A Systematic Review
Edward Goacher, Rebecca Randell, Bethany Williams
1From the Faculty of Medicine and Health (Mr Goacher and Dr Treanor) and the School of Healthcare (Dr Randell), University of Leeds, Leeds, West Yorkshire, England; and the Department of Histopathology, Leeds Teaching Hospitals, National Health Service Trust, Leeds (Ms Williams and Dr Treanor).
Context:
-Light microscopy (LM) is considered the reference standard for diagnosis in pathology. Whole slide imaging (WSI) generates digital images of cellular and tissue samples and offers multiple advantages compared with LM. Currently, WSI is not widely used for primary diagnosis. The lack of evidence regarding concordance between diagnoses rendered by WSI and LM is a significant barrier to both regulatory approval and uptake.
Objective:
-To examine the published literature on the concordance of pathologic diagnoses rendered by WSI compared with those rendered by LM.
Data Sources:
-We conducted a systematic review of studies assessing the concordance of pathologic diagnoses rendered by WSI and LM. Studies were identified following a systematic search of Medline (Medline Industries, Mundelein, Illinois), Medline in progress (Medline Industries), EMBASE (Elsevier, Amsterdam, the Netherlands), and the Cochrane Library (Wiley, London, England), between 1999 and March 2015.
Conclusions:
-Thirty-eight studies were included in the review. The mean diagnostic concordance of WSI and LM, weighted by the number of cases per study, was 92.4%. The weighted mean κ coefficient between WSI and LM was 0.75, signifying substantial agreement. Of the 30 studies quoting percentage concordance, 18 (60%) showed a concordance of 90% or greater, of which 10 (33%) showed a concordance of 95% or greater. This review found evidence to support a high level of diagnostic concordance. However, there were few studies, many were small, and they varied in quality, suggesting that further validation studies are still needed.
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Confocal Fluorescence Microscopy
Super-resolution Fluorescence Microscopy
Two-Dimensional Microscopy in Microbiology
Three-Dimensional Microscopy in Microbiology