Related Experiment Video
Updated: Mar 18, 2026

Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
Preclinical assessment of a modified Occlutech left atrial appendage closure device in a canine model
Jung-Sun Kim1, Seul-Gee Lee2, Sung-Kyung Bong2
1Severance Cardiovascular Hospital, Yonsei University College of Medicine, Seoul, Republic of Korea; Cardiovascular Product Evaluation Center.
Background:
LAA occlusion has a similar stroke prevention efficacy compared to anticoagulation treatment for non-valvular atrial fibrillation.
Objective:
The objective of this study was to assess the feasibility and safety of a modified Occlutech® left atrial appendage (LAA) closure device in a canine model.
Methods:
The device was implanted in 10 dogs (33±1kg) using fluoroscopy and transesophageal echocardiography (TEE) guidance. The modified Occlutech® LAA occlusion device was compared with the current version, the Watchman device, and the Amplazter cardiac plug (ACP). LAA occlusion and anchoring to the LAA were evaluated. All dogs were assessed using angiography, TEE, and a gross anatomy examination.
Results:
The 10 LAA occlusion devices were to be implanted into 10 dogs (5 modified Occlutech devices, 3 current version of Occlutech devices, 1 Watchman, and 1 ACP). LAA implantation was not performed in one dog due to transeptal puncture failure. The three current version of Occlutech devices were embolized immediately after implantation, so three modified devices of the same size were implanted securely without embolization. The mean implant size was 20.1±2.0mm. The devices chosen were a mean of 23.3±10.6% larger than the measured landing zone diameters. Post-implant angiography and TEE revealed well-positioned devices without pericardial effusion or impingement on surrounding structures.
Conclusions:
The results of this acute animal study suggested that a modified Occlutech® LAA occlusion device was feasible and had greater anchoring performance in canines. Additional large clinical studies are needed to evaluate safety and efficacy.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
07:41Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022