Kleihauer Testing and Flow Cytometry. A Comparative Study for Assessment of Feto-Maternal Haemorrhage
J Duguid1, I Bromilow1, J Eggington1
1a National Blood Service .; Liverpool L7 8TW.
Hematology (Amsterdam, Netherlands)
|July 14, 2016
Summary
The Kleihauer test for feto-maternal hemorrhage (FMH) shows variability. A standardized Kleihauer test is valuable for estimating FMH and guiding anti-D Immunoglobulin (Ig) dosage, with flow cytometry offering additional accuracy for complex cases.
Area of Science:
- Medical Laboratory Science
- Immunohaematology
- Obstetrics and Gynaecology
Background:
- The Kleihauer test quantifies feto-maternal hemorrhage (FMH) to determine anti-D Immunoglobulin (Ig) dosage for RhD-negative women.
- Accuracy concerns exist for the Kleihauer test, with flow cytometry proposed as a more accurate alternative.
- Flow cytometry is not universally available in all laboratories performing FMH estimations.
Purpose of the Study:
- To assess the comparability of results from standardized Kleihauer techniques with routine hospital Kleihauer tests and flow cytometry.
- To evaluate the accuracy and consistency of the Kleihauer test in quantifying FMH.
- To determine the clinical utility of a standardized Kleihauer test versus flow cytometry for guiding anti-D Ig administration.
Main Methods:
- Analysis of 957 samples comparing routine hospital Kleihauer tests with a standardized Kleihauer test and flow cytometry.
- Samples were initially tested in referring hospitals, then sent to a central transfusion center for standardized testing.
- Standardized Kleihauer technique and flow cytometric analysis were performed on all samples at the Mersey Transfusion Centre.
Main Results:
- Significant variation in Kleihauer test results was observed, even with the same sample.
- A tendency to overestimate FMH size was noted, potentially leading to unnecessary anti-D Ig administration.
- A standardized Kleihauer test demonstrated value in estimating FMH size.
Conclusions:
- A meticulously performed standardized Kleihauer test is valuable for estimating FMH and guiding anti-D Ig dosage.
- Flow cytometry provides additional value for equivocal Kleihauer results or when large FMH is indicated.
- Standardization of the Kleihauer test technique is crucial for reliable FMH quantification and appropriate clinical management.