Controlled FiO2 Therapy to Neonates by Oxygenhood in the Absence of Oxygen Analyzer
S K Jatana1, S Dhingra2, Mng Nair3
1Senior Advisor (Paediatrics), 151 Base Hospital, C/o 99 APO.
Background:
A study was conducted to evaluate a system of standardizing the oxygen concentration inside the oxygen hood and to develop guidelines for controlled FiO2 administration by changing size of the hood, lid position on the hood and the oxygen flow rate, without an oxygen analyzer. The effect of low flow rates on carbon dioxide (CO2) retention was also studied.
Method:
A dummy patient and thirty neonates, requiring oxygen to be delivered through head box, constituted the material for the study group. Oxygen content in the head box was measured using a standard oxygen analyzer while the size of head box, flow rate and lid position were changed independently and in combination. The head boxes were tested on a dummy patient. These results were analyzed, and applied to thirty neonates requiring oxygen therapy using a head box.
Result:
Volume of headbox had an inverse relation with the oxygen concentration inside the headbox and smaller sized headbox achieved more predictable oxygen concentration at all flow rates. Maximum difference in oxygen concentration by varying the lid position was seen in the large headbox. Keeping the variables constant, oxygen concentration was significantly lower in babies as compared to dummy. No significant CO2 retention was found at a flow rate of four litres per minute (lpm) in small and three lpm in a medium and large head box respectively, while lower flow rates were associated with CO2 retention.
Conclusion:
It is possible to predict the oxygen concentration inside the head box without the use of oxygen analyzer. Larger head box and higher lid position, results in lower oxygen concentration, at a given oxygen flow rate. Oxygen concentration achieved in babies is lesser than the concentration achieved in a dummy. Flow rates of less than four lpm in small and three lpm in medium and large sized head boxes are associated with CO2 retention. These results are not applicable to infants weighing less than 2 kg.
More Related Videos
07:15Preoxygenation Techniques for Tracheal Intubation in Critically Ill Adults Utilizing Oxygen Mask and Noninvasive Ventilation
Published on: December 5, 2025
05:45Delivery of In Vivo Acute Intermittent Hypoxia in Neonatal Rodents to Prime Subventricular Zone-derived Neural Progenitor Cell Cultures
Published on: November 2, 2015
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...