In vivo Structural Assessments of Ocular Disease in Rodent Models using Optical Coherence Tomography
Published on: July 24, 2020
3.6K
Posterior Eye Shape Measurement With Retinal OCT Compared to MRI
Anthony N Kuo1, Pavan K Verkicharla2, Ryan P McNabb1
1Department of Ophthalmology, Duke University Medical Center, Durham, North Carolina, United States.
Investigative Ophthalmology & Visual Science
|July 14, 2016
Summary
Optical coherence tomography (OCT) can measure posterior eye shape similarly to magnetic resonance imaging (MRI) after distortion correction. This noninvasive method may replace MRI for myopia research.
Area of Science:
- Ophthalmology
- Medical Imaging
- Biophysics
Background:
- Posterior eye shape is crucial for understanding myopia.
- Magnetic resonance imaging (MRI) is a current standard for its assessment.
- Exploring alternative, noninvasive imaging methods is essential.
Purpose of the Study:
- To evaluate optical coherence tomography (OCT) as an alternative to MRI for posterior eye shape measurement.
- To test if OCT can provide similar measurements of posterior eye shape parameters as MRI.
Main Methods:
- Analyzed macular spectral-domain OCT and brain MRI images from the Singapore Epidemiology of Eye Diseases study.
- Segmented ocular images and corrected OCT data for optical distortions.
- Fitted segmentations to spheres/ellipsoids to derive radius of curvature (Rc) and asphericity (Qxz).
- Compared MRI and OCT measurements using paired t-tests and categorical agreement analysis.
Main Results:
- Distortion correction significantly reduced the difference in posterior eye radius of curvature (Rc) between OCT and MRI (P = 0.51).
- Differences in asphericity (Qxz) between MRI and corrected OCT were not statistically significant (P = 0.28).
- Categorical agreement for prolateness/oblateness based on Qxz was moderate (κ = 0.50).
Conclusions:
- Distortion-corrected OCT measurements of posterior eye shape (Rc and Qxz) are comparable to MRI.
- Noninvasive, office-based OCT shows potential as a substitute for MRI in posterior eye shape studies.
- Further validation is needed due to moderate categorical agreement in shape classification.