Abstract:
This paper shows the opto-mechanical design of a new filter radiometer built at the Physikalisch-Technische Bundesanstalt, Germany, for the accurate determination of the thermodynamic temperature of high-temperature blackbodies. The filter radiometer is based on a three-element reflection-type trap detector that uses three large active area silicon photodiodes. Its spectral coverage and field of view are defined by a detachable narrow-band filter and a diamond-turned precision aperture, respectively. The temperature of the filter radiometer is stabilized using a water-streamed housing and is measured using a thin-film platinum thermometer placed onto the first photodiode element. The trap "mount" has been made as compact as possible, which, together with the large active area of the chosen photodiodes, allows a wide field of view. This work presents the design of the filter radiometer and discusses the criteria that have been considered in order for the filter radiometer to suit the application.