Related Experiment Video
Updated: Mar 17, 2026

A Teleoperated Robotic System-Assisted Percutaneous Transiliac-Transsacral Screw Fixation Technique
Published on: January 6, 2023
Assessment of Intra-articular Screw Penetration During Radial Head and Olecranon Locking Plate Fixation: A Cadaveric
Ismar Dizdarevic1, Claire M Eden1, Matthew Bengard1
1Mt. Sinai St. Luke's-Roosevelt Hospital, New York, NY, USA.
Background:
The purpose of this study is to evaluate the role of radiographic and clinical exams in predicting screw penetration into the proximal radioulnar joint and ulnohumeral joint during open reduction and internal fixation of the radial head and proximal ulna.
Methods:
Olecranon and radial head plates were applied to 15 cadaveric elbows. Screws were assessed for intra-articular joint penetration using both clinical exam and radiographic evaluation. Clinical exam consisted of evaluation for crepitus. Radiographs demonstrating screws positioned near the joint surface were evaluated for penetration by 3 fellowship trained hand surgeons. Elbows were disarticulated and screw prominence was determined and recorded using standardized calipers. The ability of clinical and radiographic exams to correctly predict a breach in the articular surface was determined by calculating sensitivity, specificity, and predictive values. Consideration was given to screw position.
Results:
The sensitivity of crepitus was 81.1% for screws in the radial head plate and 72.6% for screws in the olecranon plate. The sensitivity of radiographs was 72.4% for the screws in the radial head plate and 55.0% for screws in the olecranon plate. Correct radiographic assessment of penetration varied but position o-2 on the olecranon plate consistently resulted in the lowest sensitivity of 30.3%.
Conclusions:
The study evaluates sensitivity and specificity of clinical and radiographic means when assessing for articular penetration of screws during olecranon and radial head locking plate fixation. Certain screw locations are more difficult to evaluate than others and may go undetected by standard means of assessment used in a surgical setting.