Cavity-enhanced rotational Raman scattering in gases using a 20 mW near-infrared fiber laser
Abstract:
A novel cavity-enhanced laser diagnostic has been developed to perform point measurements of spontaneous rotational Raman scattering. A narrow linewidth fiber laser source (1064 nm) is frequency locked to a high-finesse cavity containing the sample gas. Intracavity powers of 22 W are generated from 3.7 mW of incident laser power, corresponding to a buildup factor of 5900. A triple monochromator and a photomultiplier tube in counting mode are used to disperse and measure the scattering spectra. The system is demonstrated with rotational Raman spectra of nitrogen, oxygen, and carbon dioxide at atmospheric pressure. The approach will allow temporally and spatially resolved Raman measurements for combustion diagnostics and, by extending to higher power, Thomson scattering for diagnostics of low-density plasmas.
Related Concept Videos
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...