Related Experiment Video
Updated: Mar 17, 2026

Laparoscopic-Assisted Seldinger Technique for Peritoneal Dialysis Catheter Insertion
Published on: May 23, 2025
A Novel Endoscopic Catheter for "Laparoscopy-Like" Irrigation and Suction: Its Research and Development Process and
Yasuhiro Miyazaki1,2, Kiyokazu Nakajima1,2, Makoto Hosaka1,3
11 Division of Next Generation Endoscopic Intervention (Project ENGINE), Global Center for Medical Engineering and Informatics, Osaka University , Osaka, Japan .
Background:
Inspired by natural orifice translumenal endoscopic surgery (NOTES), the authors launched a development of novel endoscopic irrigation and suction (I/S) catheter for "laparoscopy-like" I/S in flexible gastrointestinal (GI) endoscopy. The aims were to describe its basic research and development (R&D) process and to estimate its performance in both preclinical and clinical settings.
Methods:
In basic R&D phase, a layout of side hole at apex nozzle of endoscopic I/S (EIS) catheter were theoretically calculated and designed. Material of nozzle was selected based on the strength analysis. The performance of final prototype EIS catheter was then assessed preclinically in the porcine stomach, to compare with conventional endoscopic "tip irrigation" and "tip suction" as control. After regulatory clearance, safety and feasibility of I/S using EIS catheter were clinically assessed by endoscopists in small number of patients.
Results:
Bench tests revealed 0.4 mm in diameter, 24 holes, and 6-8 holes per circumference as most suitable layout of side holes, and polyetheretherketone as an optimal nozzle material, respectively. Time to inject 500 mL saline with the EIS catheter was significantly shorter than tip irrigation (101 ± 3.1 seconds versus 154 ± 3.1 seconds; P < .05). The EIS suction was significantly weaker than conventional endoscopic tip suction, though it remained within the practical range. No mucosal injuries were noted in the EIS suction. In clinical assessments for human use, no adverse events were observed, and high degree of satisfaction for endoscopists was obtained.
Conclusion:
The newly developed EIS catheter is safely used with satisfactory performance in flexible GI endoscopy.
Related Concept Videos
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Tracheostomy Suctioning II: Procedure
Suctioning the Nasopharyngeal Airway
Equipment Required
Urologic Endoscopic Procedure: Cystoscopic Examination
Endoscopic Procedures II: Colonoscopy