Superiority of chlorhexidine 2%/alcohol 70% wipes in decontaminating ultrasound equipment
Bhavin Shukla1, Victoria Howell1, Alicia Griffiths1
1Department of Critical Care, The Queen Elizabeth Hospital, King's Lynn, Norfolk, UK.
Abstract:
Ultrasound equipment is known to act as a reservoir for potentially pathogenic organisms. The aims of these studies were to establish current cleaning practices, to review the extent of bacterial contamination of ultrasound equipment in our hospital, to establish an effective cleaning regimen and to ensure that cleaning does not cause damage. A questionnaire was sent to all acute NHS hospitals in England to establish current cleaning practices. A review of our current practice was performed to establish the extent of bacterial contamination of ultrasound equipment currently in use. Laboratory studies compared cleaning the probes with soap and water with decontaminating with a chlorhexidine 2% and alcohol 70% wipe, including quantifying the residual effect. Accelerated aging was performed on the probe and staff surveyed to establish potential problems with using the wipes on the probe. The survey revealed that a variety of cleaning methods were used to decontaminate ultrasound probes; 57% of our ultrasound machines were contaminated with bacteria. The laboratory studies showed superiority of the chlorhexidine and alcohol wipes over soap and water due to a residual effect, both immediately after cleaning and after 24 hours. The staff survey demonstrated no apparent change in function of the probe after cleaning with the chlorhexidine wipes. Cleaning ultrasound probes with chlorhexidine and alcohol wipes is effective and provides additional protection against bacterial contamination due to its residual effect, and appears in the short term to have no detrimental effect on the probe.
More Related Videos
05:50International Expert Consensus and Recommendations for Neonatal Pneumothorax Ultrasound Diagnosis and Ultrasound-guided Thoracentesis Procedure
Published on: March 12, 2020
09:30Author Spotlight: Microbial Control and Monitoring Strategies for Cleanroom Environments and Cellular Therapies
Published on: March 17, 2023
Related Concept Videos
Hand hygiene
Hand washing...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Chemical Agents for Microbial Control
Handwashing I: Introduction and Types of Equipment
Hand wash basins in clinical areas should have faucets that can be turned on and off without using the hands; that is, they should be non-touch or lever-operated....
Handwashing II: Pre-procedure and Initial Procedure Steps
The hand washing procedure itself includes the following steps. First, cover cuts, if any, on hands with a waterproof dressing. Cuts and abrasions can become contaminated with bacteria hindering the ability to clean the area thoroughly. In addition, repeated hand washing can worsen an injury. The nails must be short and clean, without nail...