Performance of Anatomically Designed Quadripolar Left Ventricular Leads: Results from the NAVIGATE X4 Clinical Trial
Suneet Mittal1, Devi Nair2, Benzy J Padanilam3
1Valley Health System, Ridgewood, NJ, USA. mittsu@valleyhealth.com.
Introduction:
The safety and efficacy of a novel family of quadripolar left ventricular (LV) pacing leads designed to pace from nonapical regions of the LV with low pacing capture thresholds was studied in patients undergoing implantation of a cardiac resynchronization therapy defibrillator (CRT-D).
Methods And Results:
Patients receiving a CRT-D were implanted with 1 of 3 ACUITY X4 leads (Spiral Long, Spiral Short, or Straight), designed to address coronary venous anatomical variability. Electrical performance and LV lead related complications were evaluated 3 and 6 months post implantation, respectively. 764 patients (68 ± 11 years, 66% male) were enrolled; 738 (97%) successfully implanted with an ACUITY X4 lead (Spiral L, n = 239, 31%; Spiral S, n = 281, 37%; Straight, n = 218, 29%). A targeted threshold ≤2.5 V was achieved in 644 (94%) patients. The median threshold from the best proximal electrode was lower than the tip electrode (0.9 V [IQR 0.7, 1.3] vs. 1.3 V [IQR 0.7, 2.5], p< 0.001) on Spiral leads. Irrespective of lead implanted, one of the proximal electrodes was the programmed cathode in most patients. The overall LV complication-free rate was 98%. LV lead dislodgment occurred in 8 (1%) patients. PNS occurred in 58 (8%) patients, but only 3 (0.4%) patients required surgical intervention.
Conclusion:
The ACUITY X4 LV leads had low pacing thresholds particularly from proximal electrodes, a high incidence of pacing from the nondistal electrode, and low likelihood of dislodgment or PNS requiring surgical intervention. (ClinicalTrials.gov Identifier: NCT02071173).
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Cardiac Catheterization II: Right Heart Catheterization