Comparison of Routine Knee Magnetic Resonance Imaging at 3 T and 7 T
Elisabeth Springer1, Klaus Bohndorf, Vladimir Juras
1From the *High Field MR Center, Department of Biomedical Imaging and Image-Guided Therapy, and †Department of Orthopaedics, Medical University of Vienna, Vienna, Austria; ‡Siemens Healthcare Pty Ltd Australia, Imaging & Therapy Systems, Magnetic Resonance, Macquarie Park, New South Wales, Australia; §Siemens Healthcare, Erlangen, Germany; and ∥CD Laboratory for Molecular Clinical MR Imaging, Vienna, Austria.
Objective:
The aim of this study was to compare quantitative and semiquantitative parameters (signal-to-noise ratio [SNR] and diagnostic confidence) from a standard knee magnetic resonance imaging (MRI) examination with comparable sequence protocols and acquisition times at 3 T and at 7 T.
Materials And Methods:
Forty patients experiencing knee pain of unknown etiology underwent comparable MR protocols with standard turbo-spin echo and short tau inversion recovery sequences of the knee joint (5 sequences) at 3 T and 7 T. For quantitative analysis, SNR was determined using these 5 sequences and 3 additional morphological sequences. For a semiquantitative assessment of diagnostic confidence, a diagnostic confidence score (DCS) was assigned, using a 10-point scale. Two experienced radiologists who specialized in musculoskeletal imaging and who were blinded to the field-strength independently assessed 22 potential pathological findings, in total, in 4 anatomically defined areas in the knee joint and rated their diagnostic confidence.
Results:
In quantitative analysis, all sequences provided higher voxel-volume-adjusted SNR values at 7 T compared with that at 3 T. In semiquantitative analysis, summed DCS values for potential pathological findings in each of the 4 anatomically defined areas were higher at 7 T compared with that at 3 T. There was a statistically significant improvement in the DCS for both readers at 7 T for the diagnosis and exclusion of focal or diffuse grade I or II cartilage defects in the patellar cartilage. For 8 potential pathological findings, a statistically significant difference between the 2 field-strengths could be observed for 1 reader only. For the residual 13 potential pathological findings, there was no statistically significant difference observed. The percentage of concordant ratings was 84.6% at 3 T and 85.4% at 7 T.
Conclusions:
Ultra-high-field MRI at 7 T improved the overall diagnostic confidence in routine MRI of the knee joint compared with that at 3 T. This is especially true for small joint structures and subtle lesions. Higher spatial resolution was identified as the main reason for this improvement.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI