Posterior Segment Distortion in Ultra-Widefield Imaging Compared to Conventional Modalities
Ophthalmic Surgery, Lasers & Imaging Retina
|July 20, 2016
Summary
Optos ultra-widefield images show horizontal stretching distortions compared to Topcon retinal images. Combining Optos images taken at right angles can reduce these posterior segment distortions.
Area of Science:
- Ophthalmology
- Medical Imaging
- Retinal Imaging Analysis
Background:
- Optos ultra-widefield imaging (Optos 200TX) and Topcon retinal cameras are used for posterior segment imaging.
- Posterior segment distortions in Optos images require characterization compared to Topcon images.
Purpose of the Study:
- To describe and quantify posterior segment distortions in Optos ultra-widefield images.
- To compare these distortions with Topcon retinal camera images using optic disc dimensions.
- To explore a method for correcting Optos image distortions.
Main Methods:
- A comparative image analysis of 20 eyes using color fundus images from Optos and Topcon retinal cameras.
- Analysis of a model eye with grids imaged using Optos in conventional and rotated positions.
- Calculation of vertical-to-horizontal optic disc diameter ratios.
Main Results:
- Optos images exhibited a mean vertical-to-horizontal disc diameter ratio of 0.956, significantly different from Topcon's 1.083 (P < .001).
- A model eye imaged with Optos showed ratios of 0.910 (conventional) and 1.072 (rotated 90°).
- The average ratio from combined Optos images (conventional and rotated) was 0.987.
Conclusions:
- Optos ultra-widefield images demonstrate a consistent horizontal stretch distortion.
- Combining Optos images acquired at right angles effectively reduces these observed distortions.
- This method offers a way to correct for posterior segment distortions in Optos imaging.