Effect of Reconstruction Strategies for the Quantification and Diagnostic Accuracy of (123)I-FP-CIT SPECT
Related Experiment Video
Updated: Mar 17, 2026
![Semi-quantitative Assessment Using [18F]FDG Tracer in Patients with Severe Brain Injury](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F58641.jpg&w=3840&q=50)
Semi-quantitative Assessment Using [18F]FDG Tracer in Patients with Severe Brain Injury
Published on: November 9, 2018
Hiroaki Kameiyama1, Norikazu Matsutomo, Akio Nagaki
1Department of Radiological Technology, Kurashiki Central Hospital.
Purpose:
This study evaluates the effect of reconstruction strategies for the quantification and diagnostic accuracy of (123)I-FP-CIT SPECT.
Methods:
We evaluated the quantification of (123)I-FP-CIT SPECT obtained by several combinations of reconstruction using the striatal phantom. The phantom images were reconstructed using FBP and OSEM with/without attenuation correction (AC) and scatter correction (SC). We calculated the specific binding ratio (SBR) using volume of interest (VOI) analysis on each reconstructed images. For the clinical study, 40 patients who underwent (123)I-FP-CIT SPECT were selected. We grouped the patients into the normal binding group and decreased binding group according to their clinical diagnosis. The clinical images were reconstructed under the same conditions as the phantom study. The SBRs were calculated, and a receiver operating characteristic (ROC) analysis was performed to evaluate the diagnostic accuracy.
Results:
The SBRs with AC and SC significantly increased compared with no corrections. In the clinical study, although ROC analysis showed no significant difference in the all combinations of reconstruction, the area under the curve using SC and AC tended to be higher than that obtained by other reconstruction.
Conclusions:
Quantification of (123)I-FP-CIT SPECT was affected by reconstruction strategies. In addition, both the AC and SC improved the diagnostic accuracy of (123)I-FP-CIT SPECT. Our results suggest that both the AC and SC are recommended for the improving the quantification and diagnostic accuracy in (123)I-FP-CIT SPECT.
More Related Videos
06:51Utilizing 18F-FDG PET/CT Imaging and Quantitative Histology to Measure Dynamic Changes in the Glucose Metabolism in Mouse Models of Lung Cancer
Published on: July 21, 2018
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...