Coronary Computed Tomography Angiography: Patient-related factors determining image quality using a second-generation
Olivier Ghekiere1, Alain Nchimi2, Julien Djekic3
1Department of Radiology, Centre Hospitalier Chrétien (CHC), Rue de Hesbaye, 75, B-4000 Liège, Belgium; Department of Radiology, Jessa Hospital, Stadsomvaart 11, B-3500 Hasselt, Belgium; Faculty of Medicine and Life Sciences, Biomed and Reval, Hasselt University, Agoralaan, Building A and C, B-3500 Hasselt, Belgium.
Purpose:
To investigate the diagnostic confidence of Coronary Computed Tomography Angiography (CCTA) and the effect of patient-related factors on CCTA image quality using a second-generation 320-slice scanner.
Methods And Results:
200 consecutive patients (mean age 60±12years; 109 men) prospectively underwent CCTA. The mean body mass index (BMI) was 27.1±4.9kg/m(2); the median heart rate (HR) was 60.0 (interquartile range (IQR), 53.9-66.1) beats per minute (bpm). The median segment's diameter was 2.8 (IQR, 2.2-3.4) mm. For each coronary segment ≥1.5mm in diameter, two readers scored: diameter narrowing as < or ≥50%, overall diagnostic confidence and motion-related image quality, with interobserver agreement kappa-values of 0.89, 0.91 and 0.61 respectively. Seventy-nine of the 2505 evaluated segments (3.2%) had non-diagnostic image quality because of coronary calcifications (66/79; 83.5%), stent- (6/79; 7.5%), pacemaker- (2/79; 2.5%) or motion-related artifacts (5/79; 6.5%). The effect of patient-related factors on motion-related image quality was investigated by multinomial logistic regression in 181 patients with calcium score (IQR, 0-446.5). Increasing coronary diameter was the most improving image quality factor (odds ratio (OR), 1.8637; p<0.001), marginally followed by lower HR (OR, 0.9547; p<0.001) and calcium score (OR, 0.9997; p=0.04). Gender (p=0.70), age (p=0.24) and BMI (p=0.45) did not affect image quality.
Conclusion:
Using a second-generation 320-slice scanner, CCTA diagnostic confidence is predominantly affected by coronary calcifications, whereas motion-related image quality is non-diagnostic only in exceptional cases and mainly influenced by the coronary diameter. For future developments, our study findings therefore suggest greater requirements concerning spatial resolution and calcium-related artifact removal than concerning temporal resolution, especially to improve diagnostic confidence in patient groups with smaller coronary diameters.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Acute Coronary Syndrome III: Diagnostic Studies