An automatic delineation method for bone marrow absorbed dose estimation in (89)Zr PET/CT studies.
N E Makris1,2, R Boellaard3, C W Menke4
1Department of Radiology & Nuclear Medicine, VU University Medical Center, Amsterdam, the Netherlands. nikosemakris@gmail.com.
An automated method accurately estimates red bone marrow (RM) activity and absorbed dose in (89)Zr PET/CT scans. This active contour approach provides efficient and reliable results comparable to manual methods.
Area of Science:
- Nuclear Medicine
- Radiopharmaceutical Dosimetry
- Medical Imaging Analysis
Background:
- Accurate estimation of red bone marrow (RM) activity concentration and absorbed dose is crucial in (89)Zr positron emission tomography/computed tomography (PET/CT) studies.
- Manual delineation of volumes of interest (VOI) in lumbar vertebrae (LV) is time-consuming and subject to variability.
- Development of automated methods is needed for efficient and reproducible RM dosimetry.
Purpose of the Study:
- To develop and validate an automatic delineation method for estimating RM activity concentration and absorbed dose.
- To compare the accuracy and efficiency of an automated VOI method with manual delineation.
- To assess the feasibility of using an active contour approach for RM dosimetry in (89)Zr PET/CT.
Main Methods:
- Five patients with advanced colorectal cancer underwent serial (89)Zr-cetuximab PET/CT scans.
- Low-dose CT data were used to generate lumbar vertebrae (LV) VOI manually and automatically using an active contour method.
- RM activity was determined by mapping CT-derived RM VOI onto PET scans.
- Residence times and self/total RM absorbed doses were calculated using OLINDA/EXM 1.1.
Main Results:
- High correlations (r² > 0.85) were observed between manual and automated VOI methods for RM activity concentrations and total absorbed doses.
- The automated method yielded values within 5% of the manual method on average.
- The active contour-based automated method demonstrated high accuracy and efficiency.
Conclusions:
- An automated and efficient VOI method based on an active contour approach was successfully developed.
- This automated method enables accurate estimates of RM activity concentrations and total absorbed doses in (89)Zr PET/CT studies.
- The developed method offers a valuable tool for improving RM dosimetry in targeted radionuclide therapy.