Metal in Total Hip Arthroplasty: Wear Particles, Biology, and Diagnosis
Insights
Metal-on-metal total hip arthroplasty (THA) produced concerning wear debris and ion release. This led to a decline in its use due to failure mechanisms and biologic reactions.
Area of Science:
- Orthopedics
- Biomaterials Science
- Toxicology
Background:
- Metal-on-metal total hip arthroplasty (THA) was widely used, with over 600,000 procedures between 2006-2012 in the US.
- This bearing type involves articulating metal surfaces, raising concerns about wear debris and ion release.
Purpose of the Study:
- To review the mechanisms of metal wear in metal-on-metal THA.
- To examine the biologic reactions to cobalt and chromium ions released from these implants.
- To discuss the clinical presentation, diagnosis, and management of failing metal-on-metal THA.
Main Methods:
- Literature review focusing on metal wear, ion toxicology, and clinical outcomes of metal-on-metal THA.
- Discussion of diagnostic modalities including inflammatory markers, metal ion levels, imaging (radiographs, MRI with MARS), and ultrasound.
Main Results:
- Metal-on-metal articulations generate wear debris and release cobalt and chromium ions.
- These ions can elicit adverse biologic reactions, leading to implant failure.
- Clinical presentation includes pain, effusion, and pseudotumor formation.
Conclusions:
- The potential benefits of new orthopedic technologies must be weighed against novel failure modes.
- Metal-on-metal bearings exemplify the balance between innovation and clinical outcomes in THA.
- Understanding these failure mechanisms is crucial for orthopedic surgeons.
Abstract:
Total hip arthroplasty (THA) has been performed for nearly 50 years. Between 2006 and 2012, more than 600,000 metal-on-metal THA procedures were performed in the United States. This article reviews the production of metal wear debris in a metal-on-metal articulation and the interaction of cobalt and chromium ions that ultimately led to a dramatic decline in the use of metal-on-metal THA articulations. Additionally, the article reviews mechanisms of metal wear, the biologic reaction to cobalt and chromium ions, the clinical presentation of failing metal-on-metal articulations, and current diagnostic strategies. Further, the article discusses the use of inflammatory markers, metal ion levels, radiographs, metal artifact reduction sequence magnetic resonance imaging, and ultrasound for failed metal-on-metal THA procedures. When adopting new technologies, orthopedic surgeons must weigh the potential increased benefits against the possibility of new mechanisms of failure. Metal-on-metal bearings are a prime example of the give and take between innovation and clinical results, especially in the setting of an already successful procedure such as THA. [Orthopedics. 2016; 39(6):371-379.].