Reliability and validity of a smartphone pulse rate application for the assessment of resting and elevated pulse rate
Katy Mitchell1, Megan Graff2, Corbin Hedt3
1a School of Physical Therapy , Institute of Health Sciences - Houston Campus, Texas Woman's University , Houston , TX , USA.
Abstract:
Purpose/hypothesis: This study was designed to investigate the test-retest reliability, concurrent validity, and the standard error of measurement (SEm) of a pulse rate assessment application (Azumio®'s Instant Heart Rate) on both Android® and iOS® (iphone operating system) smartphones as compared to a FT7 Polar® Heart Rate monitor. Number of subjects: 111.
Materials/Methods:
Resting (sitting) pulse rate was assessed twice and then the participants were asked to complete a 1-min standing step test and then immediately re-assessed. The smartphone assessors were blinded to their measurements.
Results:
Test-retest reliability (intraclass correlation coefficient [ICC 2,1] and 95% confidence interval) for the three tools at rest (time 1/time 2): iOS® (0.76 [0.67-0.83]); Polar® (0.84 [0.78-0.89]); and Android® (0.82 [0.75-0.88]). Concurrent validity at rest time 2 (ICC 2,1) with the Polar® device: IOS® (0.92 [0.88-0.94]) and Android® (0.95 [0.92-0.96]). Concurrent validity post-exercise (time 3) (ICC) with the Polar® device: iOS® (0.90 [0.86-0.93]) and Android® (0.94 [0.91-0.96]). The SEm values for the three devices at rest: iOS® (5.77 beats per minute [BPM]), Polar® (4.56 BPM) and Android® (4.96 BPM).
Conclusions:
The Android®, iOS®, and Polar® devices showed acceptable test-retest reliability at rest and post-exercise. Both the smartphone platforms demonstrated concurrent validity with the Polar® at rest and post-exercise.
Clinical Relevance:
The Azumio® Instant Heart Rate application when used by either platform appears to be a reliable and valid tool to assess pulse rate in healthy individuals.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring pulse
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Pulse
The pulse serves as a clinical...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps: